Inter-laboratory calibration comparison of the turbine gas meter G650
Project Description
An axial turbine gas meter G650 will be used to compare low pressure test facilities over the flow range (100 ÷ 1000) m³/h. The turbine gas meter G650 will be calibrated at 8 flow rates: 1000 m³/h, 800 m³/h, 650 m³/h, 500 m³/h, 400 m³/h, 300 m³/h, 200 m³/h, 100 m³/h.
The laboratories should carry out the calibration using their normal procedures. The laboratories should use their upstream straightening pipe (at least 5x DN) and the downstream straightening pipe (at least 3x DN) during the calibration of the gas meter.
If laboratories are not able to cover all the flow rates they may calibrate the meter over part of the flow range.
Each laboratory will be responsible for the transfer of the meter to the next laboratory.
Test conditions:
Test gas: air
Test pressure: near atmospheric Test temperature: (19,5 ÷ 23,5) °C
The supposed time to make calibration and to deliver the meter to next laboratory will be approximately one week.
Final Report 2007-03-08
The project has been finished and the report can be downloaded here>>.
An axial turbine gas meter G650 was used to compare low pressure test facilities over the flow range (100 ÷ 1000) m3/h.
The turbine gas meter G650 was calibrated at 8 flow rates: 1000 m3/h, 800 m3/h, 650 m3/h, 500 m3/h, 400 m3/h, 300 m3/h, 200 m3/h, 100 m3/h.
The laboratories carried out the calibration using their normal procedures. The laboratories used their upstream straightening pipe (at least 5x DN) and the downstream straightening pipe (at least 3x DN) during the calibration of the gas meter.
If laboratories were not able to cover all the flow rates they could calibrate the meter over part of the flow range.
Test conditions:
- Test gas: air
- Test pressure: near atmospheric
- Test temperature: (19,5 ÷ 23,5) °C