Project Description

Bilateral comparison between BNM-LNE and DPLA following the EUROMET.AUV.A-K1

Procedure:
BNM-LNE is now equipped with a new reciprocity setup allowing computer driven measurements.
The method for determination of the microphone parameters (equivalent volume, resonance frequency and loss factor) are totally new.
The uncertainty budget was reviewed.

So, this bilateral comparison is necessary to validate these recent improvements.


Final Report 2004-06-17

The bilateral comparison on primary pressure calibration of LS1P and LS2aP microphones was carried out between DPLA and BNM-LNE. Two microphones from BNM-LNE have been used as transfer standards. Calibrations were performed in 20 Hz to 10 kHz range for LS1P microphones and in 31.5 Hz to 20 kHz range for LS2aP microphones.

The results for LS2aP microphone are given in the table below. The table shows the expanded uncertainties (Ua, Ub) as reported by the participants and the degree of equivalence between the two participants is given by a pair of numbers; the difference between the reported calibration results and the associated expanded uncertainty 2.(ua² ub²)½ both in dB.

Frequency
(Hz)
Expanded uncertainties in dB (k=2) BNM-LNE - DPLA
Degree of equivalence in dB
DPLA (Ua) BNM-LNE (Ub) Difference Exp. Unc.
31.5 0.080 0.045 -0.004 0.092
63 0.040 0.035 0.001 0.053
125 0.030 0.035 0.004 0.046
250 0.030 0.035 0.003 0.046
500 0.030 0.035 0.004 0.046
1000 0.030 0.035 0.002 0.046
1250 0.030 0.035 0.004 0.046
1600 0.030 0.035 0.002 0.046
2000 0.030 0.035 0.004 0.046
2500 0.030 0.035 0.005 0.046
3150 0.030 0.035 0.005 0.046
4000 0.030 0.035 0.002 0.046
5000 0.030 0.040 0.005 0.050
6300 0.030 0.040 0.013 0.050
8000 0.030 0.050 0.000 0.058
10000 0.030 0.065 0.008 0.072
12500 0.040 0.085 0.000 0.094
16000 0.050 0.120 0.017 0.130
20000 0.080 0.200 -0.008 0.215

The results for LS1P microphone are following:

Frequency
(Hz)
Expanded uncertainties in dB (k=2) BNM-LNE - DPLA
Degree of equivalence in dB
DPLA (Ua) BNM-LNE (Ub) Difference Exp. Unc.
20 0.060 0.035 -0.011 0.069
25 0.050 0.035 -0.009 0.061
31.5 0.050 0.030 -0.008 0.058
63 0.030 0.030 -0.004 0.042
125 0.025 0.030 0.000 0.039
250 0.025 0.030 0.001 0.039
500 0.025 0.030 0.004 0.039
1000 0.025 0.030 0.005 0.039
1250 0.025 0.030 0.003 0.039
1600 0.025 0.030 -0.002 0.039
2000 0.025 0.030 0.001 0.039
2500 0.030 0.035 0.004 0.046
3150 0.030 0.040 0.002 0.050
4000 0.030 0.045 -0.010 0.054
5000 0.050 0.055 -0.009 0.074
6300 0.050 0.075 -0.023 0.090
8000 0.050 0.090 0.037 0.103
10000 0.080 0.150 -0.074 0.170

Degree of equivalence between both partners shows that BNM-LNE primary pressure microphones calibration system is in reasonably good agreement with DPLA. The observed differences in the calibration results are all within the expanded uncertainties.

Subjects
Acoustics, Ultrasound and Vibration (AUV)
Coordinator
Jean-Noël Durocher, LNE (France)
Coordinating Institute
LNE (France)
Participating Partners
Further Partners
DPLA (Denmark)

Information

Reg. No.
805
Collaboration Type
Comparison
Status
completed
Starting
2003-05-14
Completion
2004-06-14