EURAMET DOSEtrace supplementary comparison
Project Description
This comparison will compare calibration coefficients in terms of H*(10) for photon radiation.A transfer chamber will be used. Five radiation qualities are selected - N-40, N-100, N-200, S-Cs and S-Co.
Comparison funding is provided by the Project EMPIR 17RPT01 DOSEtrace "Research capabilities for radiation protection dosimeters".
The technical protocol can be downloaded here>>
Final Report 2023-06-30
The comparison has been completed and results are available in the KCDB.
.Progress Report 2022-01-18
All measurements were finished in July 2021. Draft A was created in September 2021 and sent to partners for comments. Draft B was completed in November 2021 and submitted to TC IR. Draft B is currently under review in TC IR.
Progress Report 2020-03-26
Comparison is currently halted, due to coronavirus outbreak in Europe. Until the outbreak, 2 blocks out of three were finished, and third quality control was scheduled next. As of today, we still think that the comparison can be finished in time.