Pressure Measurements and Calculation of the Effective Area of a Transfer Piston-Cylinder Assembly from 0.05 to 1 MPa

Project Description

A piston cylinder assembly of about 10 cm² area is circulated between the participating laboratories. Its area at 20°C, for the actual pressure, is determined by the different laboratories from 0.05 Mpa up to 1 MPa. The measurement is made in gauge mode with N2 as working fluid.


Final Report 2006-05-30

The project has been completed and the results are available in the KCDB.

The pressure realised at CMI and at METAS has been compared using a piston-cylinder of 10 cm2 as artifact. The measurements show the equivalence of the pressure defined by the two laboratories. Dimensional measurement of the artifact is in agreement with the values obtained by cross-floating.

 

Subjects
Mass and Related Quantities (M)
Coordinating Institute
METAS (Switzerland)
Participating Partners
CMI (Czechia)

Information

Reg. No.
788
Registered in KCDB
Collaboration Type
Comparison
Status
completed
Starting
2004-03-20
Completion
2006-05-30